Step by Step Guide for Gas Flow Meter Calibration

 gas flow meter is used to measure the volumetric or mass flow rate of various process gases in industrial plants and machinery. The flow rate refers to the speed at which fluid is moving through closed pipes or ducts at a given time.

Modern instrumentation & control engineering design needs monitoring of fluid flow rates to optimize the use of gases and improve the efficiency of processes and equipment. Therefore, a gas flow meter device must be calibrated against a reference or a master calibration system.


Due to process & environmental conditions flowmeters may have some drift in flowrate readings which needs recalibration periodically for the flow range it is designed for, to achieve the desired accuracy for smooth process operation.


Some organizations have an ISO audit system in place which needs recalibration at a specific time frame.


Types of Gas Flow Meters Available in Market

Various flowmeter technologies are available currently in the market which needs to be calibrated and there are different methods for calibration based on product design. 


  • Turbine flowmeter
  • Vortex flowmeter
  • Coriolis flowmeter
  • Thermal Mass flowmeter
  • Ultrasonic flowmeter

and there are different methods for calibration based on product design.  


Types of Calibration Procedures

Gas flowmeter calibration systems mainly depend on product technology to be calibrated.


Some widely used calibration procedures are as below


  •  Reference Master Meter calibration
  • Primary Air by Bell / Piston prover calibration
  • Critical Flow Venturi (Sonic nozzle) calibration
  • Wind Tunnel Air Velocity calibration
  • Onsite validation by calibrated reference gas flowmeter

Gas flow meter calibration is done as per above various calibration systems which involve comparison and adjustment of the flowmeter under test confirming to National & International standards mainly following ISO 17025 standards for calibration methods.


Reference Master Meter Calibration Procedure

Reference Master meter calibration compares with duly calibrated master gas flowmeter traceable to national accredited calibration laboratories with a gas flowmeter under test at desired flow ranges and adjusts its calibration as per desired accuracy tolerances.


Reference Master meter calibration steps


  • Place the reference master meter in series with the flow meter under test.
  • Compare the readings of the master flow meter and flow meter using air or standard gas flow rate with a close loop calibrated volumetric tank storage system.
  • Calibrate the flowmeter under test to conform with the master flow meter calibration.

Piston / Bell Prover Gravimetric Calibration Procedure

Gravimetric calibration is one of the most accurate and cost-effective volumetric and gas mass flow meter calibration procedures at low pressure.

Piston prover-based gravimetric method is ideal for gases for low flow rates whereas Bell Prover calibration is mainly used for a high flow rate of air calibration.

The calibration flow laboratory is always maintained at controlled ambient conditions such as temperature, humidity for master flow meters’ best performance.

In the piston or bell prover flow meter calibration procedure, a known volume of fluid is forced through a flow meter under test. The piston/bell prover is a cylindrical device with a known internal diameter.

The piston/bell prover contains a piston/ bell which produces volumetric flow by positive displacement.

The piston prover method is ideal for Low flow rate gas or liquid thermal mass flow controller, ultrasonic flow meter & turbine flow meter calibration to a high degree of accuracy.

Piston / Bell prover Gravimetric calibration steps

  • Fix the flowmeter in a series of piston / bell prover systems for calibration.
  • Desired gas flow via test flowmeter enters in the piston / bell prover which will move piston / bell upwards with detecting distance between sensors for known volume between is measured.   
  • Measuring travel time, pressure, temperature and flow rate at reference conditions is calculated via software systems.
  • Compare the calculated flow rate with the flowmeter under test and adjust it to the measured flow rate.

Critical Flow Venturi (Sonic Nozzle) Calibration Procedure
Critical Flow Venturi Nozzle (Sonic Nozzles) is used for different flow rate capacities designed as per ISO 9300:2015 as a reference standard.

Sonic nozzles are considered the best reference standard for calibration of precision flow meters used for custody transfer.

Gas flow accelerates to the critical velocity at the throat (known as sonic velocity) where it will reach a steady flow state of single-phase gases.

This sonic velocity is a specific point velocity at a certain pressure which is accurate as a reference velocity for calibrating a flowmeter under test. It is used for almost all fixed type flowmeters mainly vortex flowmeter, turbine flowmeter, thermal mass flowmeter, Coriolis flowmeter, etc.   

Sonic nozzle calibration steps

  • Fix the flowmeter in series of the pipeline where calibrated sonic nozzle
  • Start air compressor in the testing system
  • When sonic velocity reaches the measuring section, the flowmeter under calibration needs to adjust its reading & set it. Repeat the same steps for different sonic velocity points for specific flow rate.
  • Comparing flow rate from sonic nozzle with flow meter under calibration.

Wind Tunnel Calibration Procedure

The wind tunnel is designed for characterization of air velocity, air turbulence for various different turbomachinery, and instruments calibration and simulation.


There are different proprietary & standard designs of wind tunnels installed. It has an open test chamber for calibration or test of the device under test.


It is widely used for various types of flow sensor calibration such as hot-wire/vane anemometer, insertion thermal mass flowmeter, pitot tube, measuring probes, etc.  


Wind tunnel calibration steps


  • Install a flow sensor probe in the test chamber with desired insertion depth and correct orientation.
  • Step by step increasing reference velocity and measure & record flow sensor under test.
  • Based on reference velocity, plot a curve and suitable interpolation & extrapolation flow calculations upload curve into device under test.

Onsite Validation & Calibration Procedure
Onsite validation and calibration of gas flow meters are also possible in some industrial applications. This method is not accurate as of the above methods but can validate gas flowmeters in use having high errors or non-repeatability.


 
Mainly two ways are possible for onsite validation

  • Pump-up test with pressure vessel
  • With calibrated reference gas flowmeter to be installed in series
Pump-up test steps

  • Pressure vessel in series of gas flowmeter
  • Start Gas flow and note gas flow meter reading and timer for filling pressure vessel
  • Pressure vessel pressure with respect to volume capacity and gas flow meter totaliser to be compared
Validation is done by comparing both results and errors. 
  • Calibrated reference gas flowmeter

  • Install calibrated reference gas flowmeter in series of gas flowmeter under validation
  • Compare both flowmeter result and compute error

Comments

  1. Thanks for suggesting good list. I appreciate your work this is really helpful for everyone. Get more information at astm a193 grade b8. Keep posting such useful information.

    ReplyDelete
  2. I agree with a lot of the points you made in this article. If you are looking for the Stainless Steel 304 round bar suppliers, then visit nickelalloysonline. I love your content, they are very nice and very useful to us and this text is worth everyone’s attention.

    ReplyDelete
  3. This is a great article with lots of informative resources. I appreciate your work this is really helpful for everyone. Check out our website finned tubes manufacturer in india for more vishalsteelindia related info!

    ReplyDelete
  4. I found your blog on Google and read a few of your other posts. I just added you to my Google News Reader. You can also visit ASTM A193 B6 Bolts for more Viha Steel & Forging related information and knowledge, Keep up the great work Look forward to reading more from you in the future.

    ReplyDelete
  5. Information is very informative also you can click structural steel plate suppliers ,and get such type of info, this is the great resource to get such type of information.

    ReplyDelete
  6. Wow, excellent post. I'd like to draft like this too - taking time and extremely hard work to make a great article. Visit stud bolt manufacturers in india. This post has inspired me to write some posts that I am going to write soon.

    ReplyDelete
  7. Really helpful down to the ground, happy to read such a useful post. I got a lot of information through it and I will surely keep it in my mind. Keep sharing. If you are looking for some useful data and information regarding stainless steel butterfly valve manufacturers in india then visit Rache process equipment LLP.

    ReplyDelete
  8. I have gone through many blogs but yours puts a great knowledge inside and I completely satisfied with it, hoping more Information like this. Well, meanwhile I would like to share about Carbon steel pipe fittings manufacturers. You can also visit for more related information.

    ReplyDelete

Post a Comment

Popular posts from this blog

PLC Program for Mixing Tank

 Create a ladder diagram for controlling a batch mixing process. Implement a PLC program for mixing tank or Mixing Process using PLC Ladder Logic. PLC Program for Mixing Tank Fig : Mixing tank A tank is used to mix two liquids. The required control circuit operates as follows: A. When the START button is pressed, solenoids A and B energize. This permits the two liquids to begin filling the tank. B. When the tank is filled, the float switch trips. This de-energizes solenoids A and B and starts the motor used to mix the liquids together. C. The motor is permitted to run for 1 minute. After 1 minute has elapsed, the motor turns off and solenoid C energizes to drain the tank. D. When the tank is empty, the float switch de- energizes solenoid C. E. A STOP button can be used to stop the process at any point. F. If the motor becomes overloaded, the action of the entire circuit will stop. G. Once the circuit has been energized, it will continue to operate until it is manually stopped. Solution...

What is Relay? How it Works? Types, Applications, Testing

 We use relays for a wide range of applications such as home automation, cars and bikes (automobiles), industrial applications, DIY Projects, test and measurement equipment, and many more. But what is Relay? How a Relay Works? What are the Applications of Relays? Let us explore more about relays in this guide. What is a Relay? A Relay is a simple electromechanical switch. While we use normal switches to close or open a circuit manually, a Relay is also a switch that connects or disconnects two circuits. But instead of a manual operation, a relay uses an electrical signal to control an electromagnet, which in turn connects or disconnects another circuit. Relays can be of different types like electromechanical, solid state. Electromechanical relays are frequently used. Let us see the internal parts of this relay before knowing about it working. Although many different types of relay were present, their working is same. Every electromechanical relay consists of an consists of an Elect...

Chlorine dioxide Analyzer Principle

 Chlorine dioxide measurement Chlorine dioxide (ClO2) is an instable, non-storable, toxic gas with a characteristic scent. The molecule consists of one chlorine atom and two oxygen atoms – represented in the chemical formula ClO2. It is very reactive. To avoid the risk of spontaneous explosions of gaseous chlorine dioxide or concentrated solutions, it is generally handled in dilution with low concentrations. ClO2 is soluble in water, but tends to evaporate quickly. Typically it is prepared on site, for example from hydrochloric acid and sodium chlorite. The procedure provides solutions with approx. 2 g/l ClO2 that can be safely handled and stored for several days. Image Credits : krohne Sensor Parts : Reference electrode Applied chlorine dioxide specific potential Current needed to maintain the constant potential Counter electrode Measuring electrode The disinfection effect of ClO2 is due to the transfer of oxygen instead of chlorine, so that no chlorinated byproducts are formed. C...