In today’s environment of shrinking cost margins and increased regulations, maintaining properly calibrated equipment becomes ever important. Even the smallest inaccuracy in process measurements can cause a significant loss in revenue, especially if the process handles large volumes.
Electronic equipment can lose accuracy over a period of time, leading to incorrect readings; this is where calibration comes in. Calibrating electronic equipment, in this case transmitters, brings the measured values of the device in line with a known value from an applied standard.
Instrument Loops
In order to understand transmitter calibration, it may be beneficial to first review instrument loops. An instrument loop refers to a configuration of equipment, connected in such a way to deliver power to all devices and obtain a reading from these devices. An instrument loop can be either electrical or pneumatic.
The figure below shows a simple instrument loop with a power supply, transmitter, and a meter to measure current in the loop.
- Some commonly used constants are below:
- “Standard/International” gravity is 980.665 cm/sec2
- US mean gravity is 980.000 cm/sec2
- All instrument calibrations should be made to .1% of calibrated span.
- Isolate the transmitter from the source. This means taking the transmitter off service.
- Connect a primary device, such as a dead weight tester or a decade box, to the input of the transmitter. Make sure the primary device has been certified. The setup should be similar to the figure below. In this figure, a HART communicator has also been connected to aid in programming a smart transmitter.
- Record on the appropriate form. A sample chart is shown below for a 0 to 150“ differential pressure transmitter.
- Apply the desired input from the table to the transmitter and read the output. Record it in the chart under the “As Found” column.
- If the “As Found” number is outside of the minimum acceptable and the maximum acceptable ranges, then the transmitter must be recalibrated. Even if the readings are inside the limits, but still a bit off, it is a good idea to recalibrate the transmitter.
- Recalibrate the transmitter according to the procedure for that transmitter. This usually involves putting the 0% value of input on the transmitter and zeroing the transmitter to 4 milliamps out. Then, put the 100% value on the transmitter and set the span to read 20 milliamps.
- Perform the calibration checks again as in step 4 and record the values in the “As Left” column.
Comments
Post a Comment