Skip to main content

Calibration of Gas Detectors

 A gas detector is a device that detects the presence of gases in an area, often as part of a safety system. This type of equipment is used to detect a gas leak and interface with a control system so a process can be automatically shut down. A gas detector can sound an alarm to operators in the area where the leak is occurring, giving them the opportunity to leave. This type of device is important because there are many gases that can be harmful to organic life, such as humans or animals.


Gas detectors can be used to detect combustible, flammable and toxic gases, and oxygen depletion. This type of device is used widely in industry and can be found in locations, such as on oil rigs, to monitor manufacture processes and emerging technologies such as photovoltaic. They may be used in firefighting.


Gas leak detection is the process of identifying potentially hazardous gas leaks by sensors. These sensors usually employ an audible alarm to alert people when a dangerous gas has been detected. Common sensors include infrared point sensors, ultrasonic sensors, electrochemical gas sensors, and semiconductor sensors. More recently, infrared imaging sensors have come into use. All of these sensors are used for a wide range of applications and can be found in industrial plants, refineries, waste-water treatment facilities, vehicles, and homes.


Calibration of Gas Detectors:  SC111/112 General Monitors


When Gas detector Model SC111/112 General Monitors power-up, it will take an initial power-up period of 1 minute approximately, and is observed by the SC111 during which it display “PU”. This is to allow the sensor to stabilize, then the display should read ‘0’ if there is no gas present at the sensor. If doesn’t occur then refer to manual book, in section troubleshoot.

Calibration of SC111/112 General Monitors:

Ensure that the SC111 has stabilized for at least 1 hour, and there is no combustible gas present at the sensor. A true zero reading will be obtained when the reading stabilized at the lower value.

Place the magnet over the General Monitors logo at the surface body. ‘–‘ will appear on the display first to indicate that the magnet has been positioned correctly. Then the display will begin to flash. After a total of 9 second ‘AC’ will be displayed, indicating that the unit is in the auto-calibration, then remove the magnet.

Use a General Monitors Portable Purge or Calibration Chamber to apply gas at 50% LEL (+/-5%) to the sensor. When the SC111 detects this gas it will display ‘CP’ , this mean ‘CP’ = CALIBRATION IN PROGRESS

Wait until ‘CC’ is displayed before removing gas. This will normally take less than 2 minutes, ‘CC’ = CALIBRATION COMPLETE

When the gas disperses from the sensor the SC111 will leave CALIBRATION MODE and return to a normal monitoring condition. The display should read ‘0’ when the gas has dispersed.

If the above does not occur as describe and a different code is displayed..go to the TROUBLESHOOTING section.

Calibration Check on SC111/112 General Monitors

After perform calibration on Gas detector SC111/112 then will continue to perform calibration check:

Position the magnet over the General Monitors logo, The symbol ‘–‘ will be displayed for three second and then will begin to flash

Remove the magnet and the display will now flash the gas concentration at the sensor. The analog output will be held at 1.5mA regardless of the gas concentration at the sensor.

If gas is not applied within 6 minutes the analog output will fall to 0mA and the display will read ‘F2’. To recover from this position, replace the magnet over the General Monitors Logo, repeat step 1 and 2 then proceed to step 4 within the timeout period

Apply gas at 50% LEL to the sensor. Observe that the gas reading settles at 50+/-5%. Should the final response fall outside this limit, a full calibration is required.

Note : The sensor should be exposed to clean air conditions for at least two minutes prior to entering calibration mode.

The display will continued to flash and the analogue output will remain at 1.5mA until the gas has been removed and the level at the sensor drops below 3% LEL approx. Normal monitoring will then be resumed (i.e the display will give a steady reading and the analogue output will follow the gas concentration at the sensor)

Calibration of Gas Detectors: BW Killark Gas Point


Calibration of BW Gas Point can be executed at any time during normal operation except the self test period (from 10 minutes before the self test until the self test is complete) BW technologies recommends a premium grade calibration gas, Gases with NIST (National Institute of standards and Technology)


Start Calibration:


Press and hold the external button down while the LCD displays the alarm setpoints, continue to hold the button until the display reads CAL (approx 5 seconds) and then release the button.

– The 4-20 mA output will be 3 mA throughout calibration. Calibration the Gas Point will not cause false alarms at the controller


– on this step: first LCD monitor will display low and high alarm (approx 8 sec), next the CAL icon lights for 3 sec, gas type is constantly displayed and backlight is activated.


Auto Zero:


The Gas Point will then take a zero level reading, combustible and toxic sensors; if background gas is present, apply zero the sensor, restart the calibration sequence will take 30 to 60 seconds.

– On display, numeric display will read 00, auto zero advice flashes, gas monitor constantly displayed.


Auto Span:


Insert cal cap and apply gas to sensor for approx 2 minutes (ammonia 5 minutes)

When the countdown (300 to 00) begins, span is complete, disconnect the gas

– If span fails: check calibration gas  cylinder used and concentration expected. Replace the cylinder and/or change the cal gas expected value, if required, re-calibrate


– Oxygen sensor, use pure air calibration gas (20.9% O2) in case of deficient to enriched atmosphere.


– on display : numeric display will show calibration gas value expected, gas cylinder icon flashes, span advice lights, gas type is constantly displayed, after a successful calibration, gas point automatically returns to normal operation and displays the current reading (ppm or %) present.


Note: Gas Detectors Calibration Procedure may change from vendor to vendor




Comments

  1. Pretty good post. I just stumbled upon your blog and wanted to say that I have really enjoyed reading your blog posts. I hope you post again soon. Big thanks for the useful info. smoke detector fire alarm system

    ReplyDelete
  2. Thanks for giving such good knowledge. I got some helpful information from this post. Keep Posting such types of posts. Gas Detectors Supplier in Saudi Arabia

    ReplyDelete
  3. I am attracted by the presentation of this article. It is a genuinely a gainful article for us. Keep posting, Thank you.Duevi Outdoor Wired and wireless Motion Detectors Lebanon

    ReplyDelete

Post a Comment

Popular posts from this blog

PLC Program for Mixing Tank

 Create a ladder diagram for controlling a batch mixing process. Implement a PLC program for mixing tank or Mixing Process using PLC Ladder Logic. PLC Program for Mixing Tank Fig : Mixing tank A tank is used to mix two liquids. The required control circuit operates as follows: A. When the START button is pressed, solenoids A and B energize. This permits the two liquids to begin filling the tank. B. When the tank is filled, the float switch trips. This de-energizes solenoids A and B and starts the motor used to mix the liquids together. C. The motor is permitted to run for 1 minute. After 1 minute has elapsed, the motor turns off and solenoid C energizes to drain the tank. D. When the tank is empty, the float switch de- energizes solenoid C. E. A STOP button can be used to stop the process at any point. F. If the motor becomes overloaded, the action of the entire circuit will stop. G. Once the circuit has been energized, it will continue to operate until it is manually stopped. Solution...

What is Relay? How it Works? Types, Applications, Testing

 We use relays for a wide range of applications such as home automation, cars and bikes (automobiles), industrial applications, DIY Projects, test and measurement equipment, and many more. But what is Relay? How a Relay Works? What are the Applications of Relays? Let us explore more about relays in this guide. What is a Relay? A Relay is a simple electromechanical switch. While we use normal switches to close or open a circuit manually, a Relay is also a switch that connects or disconnects two circuits. But instead of a manual operation, a relay uses an electrical signal to control an electromagnet, which in turn connects or disconnects another circuit. Relays can be of different types like electromechanical, solid state. Electromechanical relays are frequently used. Let us see the internal parts of this relay before knowing about it working. Although many different types of relay were present, their working is same. Every electromechanical relay consists of an consists of an Elect...

Ferrules and Cross Ferruling

 Ferrules are identification labels provided for every wire terminations in an instrument, equipment, or electrical/instrumentation control panels. These tube-shaped sleeves can be inserted easily on each individual wire in a multi-core cable. In earlier days fixed digits/letters are used as ferrules, but now Instrumentation engineers/technicians prints out desired ferrules by using a ferrule printing machine. Typical Ferrule The numbers/ letters on the ferrules will be given as per the approved electrical hook up or loop diagrams. This helps technicians to easily identify a particular loop/wiring from a series of terminal blocks and to troubleshoot the desired terminal connection. Separate numbers on the ferrules distinguish the positive and negative polarities of wires, thus ensure the polarity protection of the instrument. Cross Ferruling  As a wire is connected on its both ends, it is quite useful to use a cross reference method for wire identification. Unlike normal ferru...