pH Analyzer Calibration and Troubleshooting

Calibration of the pH analyzer usually concerns itself with the calibration of the probe as the electronics of the transmitter are inherently stable.

Before calibrating the probe it is critical that the probe is clean and in good condition with the proper buffer solution (see troubleshooting and maintenance section).

pH Analyzer 


Observe these practices when calibrating the pH analyzer:

  • Always rinse the probe with distilled water between calibration points.
  • Never “wipe” the probe to remove excess fluid.
  • Always ensure that the buffer solution inside the probe is a minimum 2/3 full.
  • Keep the probe in a vertical position whenever possible.
There are two main calibration techniques:
  1. One-Point Calibration
  2. Two Point calibration
One-Point Calibration
  • Ensure the temperature setting on the meter corresponds to the temperature of the buffer solutions.
  • Ensure the probe is clean and rinse it with distilled water.
  • Place the probe in a buffer solution of 7.00 pH.
  • Adjust the zero (or equivalent) setting so the device reads 7.00.
  • Rinse the probe with distilled water and return to service.
Two-Point Calibration
  • Ensure the temperature setting on the meter corresponds to the temperature of the buffer solutions.
  • Ensure the probe is clean and rinse it with distilled water.
  • Place the probe in a buffer solution of 7.00 pH.
  • Adjust the zero (or equivalent) setting so the device reads 7.00.
  • Place the probe in buffer solution 4.00 pH or 10.00 pH (depending on whether the process fluid is an acid or base.)
  • Adjust the gain (or equivalent) setting for 4.00 or 10.00 pH.
  • Rinse the probe with distilled water and return to service.
Troubleshooting the pH Analyzer
90% of all problems with a pH measuring system are problems with the probe itself.

If there is no output signal or a maximum output signal from the transmitter, check the:

  • Loop continuity.
  • Loop power supply.
  • The transmitter itself and follow the manufacturer’s electronic troubleshooting guide.

Cleaning and checking the probe at regular intervals (dependent on the process) is important. Depending on process conditions, the probe should be checked at regular intervals for proper levels of buffer solution.

The light coating of the probe can be cleaned by rinsing with a water jet or placing the probe under a faucet of running water, always rinse with distilled water.

Heavier coatings may require a rinse or brush with a mild acid or detergent solution. It may be necessary to soak the probe for a period of time to loosen the coating. Always rinse with distilled water.

Major coatings may require brushing or soak with a 2%-10% hydrochloric acid (HCl) solution. Always rinse with distilled water. Check for proper temperature compensation.

If the reading does not match what the pH value of the process is:

  • Clean the probe as per the preceding procedure.
  • The reference electrode buffer solution may be contaminated, clean the probe, replace the solution, and calibrate the system.
  • The buffer solution in the measurement or reference electrode may be depleted, fill to the required level.
  • The silver/silver chloride wire of the measurement probe may have been chemically attacked by the process fluid, change the probe.
  • Natural aging of the measurement electrode (they have a finite life span), change the probe.

Troubleshooting Guide:

Most manufacturers approved a troubleshooting guide, follow the guide to determine the cause and solution, a typical guide is shown below.



Comments

Popular posts from this blog

PLC Program for Mixing Tank

 Create a ladder diagram for controlling a batch mixing process. Implement a PLC program for mixing tank or Mixing Process using PLC Ladder Logic. PLC Program for Mixing Tank Fig : Mixing tank A tank is used to mix two liquids. The required control circuit operates as follows: A. When the START button is pressed, solenoids A and B energize. This permits the two liquids to begin filling the tank. B. When the tank is filled, the float switch trips. This de-energizes solenoids A and B and starts the motor used to mix the liquids together. C. The motor is permitted to run for 1 minute. After 1 minute has elapsed, the motor turns off and solenoid C energizes to drain the tank. D. When the tank is empty, the float switch de- energizes solenoid C. E. A STOP button can be used to stop the process at any point. F. If the motor becomes overloaded, the action of the entire circuit will stop. G. Once the circuit has been energized, it will continue to operate until it is manually stopped. Solution : A

What is Relay? How it Works? Types, Applications, Testing

 We use relays for a wide range of applications such as home automation, cars and bikes (automobiles), industrial applications, DIY Projects, test and measurement equipment, and many more. But what is Relay? How a Relay Works? What are the Applications of Relays? Let us explore more about relays in this guide. What is a Relay? A Relay is a simple electromechanical switch. While we use normal switches to close or open a circuit manually, a Relay is also a switch that connects or disconnects two circuits. But instead of a manual operation, a relay uses an electrical signal to control an electromagnet, which in turn connects or disconnects another circuit. Relays can be of different types like electromechanical, solid state. Electromechanical relays are frequently used. Let us see the internal parts of this relay before knowing about it working. Although many different types of relay were present, their working is same. Every electromechanical relay consists of an consists of an Electroma

Chlorine dioxide Analyzer Principle

 Chlorine dioxide measurement Chlorine dioxide (ClO2) is an instable, non-storable, toxic gas with a characteristic scent. The molecule consists of one chlorine atom and two oxygen atoms – represented in the chemical formula ClO2. It is very reactive. To avoid the risk of spontaneous explosions of gaseous chlorine dioxide or concentrated solutions, it is generally handled in dilution with low concentrations. ClO2 is soluble in water, but tends to evaporate quickly. Typically it is prepared on site, for example from hydrochloric acid and sodium chlorite. The procedure provides solutions with approx. 2 g/l ClO2 that can be safely handled and stored for several days. Image Credits : krohne Sensor Parts : Reference electrode Applied chlorine dioxide specific potential Current needed to maintain the constant potential Counter electrode Measuring electrode The disinfection effect of ClO2 is due to the transfer of oxygen instead of chlorine, so that no chlorinated byproducts are formed. ClO2