Two rotating impellers, designed with a figure eight cross section, rotate in opposite directions due to the forces exerted by the gas flow being measured. The shape of the impellers prevents contact while the gap between them remains constant.
Lobed Impeller Flow Meters
A gear drive external to the measuring chamber synchronizes the impellers. During each rotation four crescent shaped volumes are moved through the measuring chamber. The number of rotations is proportional to the flow. The rotation is coupled using an adjustable fine tooth gear train to the totalizer.
An unmeasured flow, which is a function of the pressure drop, flows through the gaps. The negative error is compensated by an adjustment. The viscosity of flow increases at high pressures and reduces the losses in the gaps which compensates for the higher losses which would otherwise exist due to the higher pressure difference.
The pulsations in the flow discharge can cause the pipe system connected to the meter to vibrate. If resonance should occur, loud noises and sudden pressure drops can result. This condition should not be allowed to occur; if necessary noise or pulsation dampers should be utilized.